13C Infrared Spectrometer
13C Urea Breath Test Kit
13C Methacetin Breath Test Kit
DDG Analyzer System
Heater Probe System (P)
Heater Probe System (U)
Heater Probe System (D)
13C Infrared Spectrometer
DDG Analyzer System
 
 


13C Infrared Spectrometer

Product Type Producing Area
HG-IRIS200 China
     HG-IRIS200 infrared spectrometer is high precious device for detection of 13CO2 /12CO2 ratio in the breath test. the relative error is lower than 0.3δ. HG-IRIS200 have the same veracity and accuracy with mass spectrometer. HG-IRIS200 infrared spectrometer have high stability , reliability and best performance/price ratio.
The Theory Of 13C Infrared Spectrometer
        HG-IRIS200 13C infrared spectrometer is composed of high selective 13CO2 and 12CO2 detector,which can detect CO2 between 2-8μm wave length according to the special absorbing summit of CO2 in the infrared spectrum.
        HG-IRIS200 13C infrared spectrometer have patented special optical path and biophysical design, which decrease interference between 13CO2 and guarantee the higher accuracy and veracity of the system. For sake of the accuracy of the system the thermostat have been used for temperature stability.
Features Basic data
    ◆ No radiation, high accuracy , non-invasive, operated briefly and quickly.
    ◆ High accuracy: the relative error lower than 0.3δ of 13C.
    ◆ Operated simply:collect expiratory sample of the patients and start the device by the test finished and the results analyzed and printed automatically.
    ◆ Test Speed: one sample test need less than 5 min. Ten channels work simultaneously.
    ◆ Strong data management: the software is available for storage, search, sort and statistical analysis.
    ◆ The device have ability of automatic self-test and remote diagnosis: the users can understand the device quickly and simply because of the unique self-test , self-diagnosis, self-calibration , self-restoration and internet remote diagnosis.
    ◆ Low cost operation
    ◆ Intuitive operative interface with multi- language option
         ◆ Dimension of The Device: 65×55×45 cm
         ◆ Weight: 25kg
         ◆ Main Power: 220V 50Hz
         ◆ Power Input: 300VA
         ◆ Concentration range of the sample can be              tested: CO2 concentration 1~10%
         ◆ Sample Volume: no less than 120ml/bag
Operation Conditions
         ◆ Ambient Temperature: 10℃~30℃
         ◆ Relative Humidity: ≤70%
         ◆ Atmospheric Pressure: 70kPa~106kPa
         ◆ Power Supply: 220V±22V,50Hz±1Hz
Capability
         ◆ Degree of Precision﹤0.3‰δ13C
         ◆ Analyses Speed ﹤5min/ sample
Advantages For Clinical Application
         ◆ No Radiation         ◆ Precision
         ◆ Non-invasive         ◆ Quickly
The Clinical Application Of 13C Breath Test Dependability Of Results Between Infrared Spectrometer And Mass Spectroscopy
    ◆Gastrointestinal tract
    1. Diagnosis and clinical studies for Helicobacter infection
    2. Test for bacterial overgrowth in the jejunum
    3. Measurement of gastric emptying function of solid food
    4. Test for colon fermentation and emptying time of appendix
    5. Measurement of gastric emptying function of fluid food or semifluid food
    6. Test for the activity of lactase and invertase
    ◆ Liver, gall bladder or pancreas
    1. Test for liver reserve function
    2. Exocrine pancreas function test, especially for monitoring the effect of enzyme replacement treatment
    3. Diagnosis for alcoholic cirrhosis
    4. Test for activity of pancreas lipidase and the malabsorption of lipid
    5. Enzymoprivic diagnosis for acetaldehyde dehydrogenase
    6. Measurement for enterohepatic circulation of bile acid
    ◆ Others
    1. Epilepsy
    2. Test for myocardial metabolism function
    3. Test for aminoasids metabolism function